Environmental & Scientific Instruments Co.
Environmental & Scientific Instruments Co.

 


Contact Us

Plot – 123, Industrial Area, Phase – II, PANCHKULA, Haryana - 134113, India

Mobile : +91-9216626616

Phone : +91-172-5033251

Fax : +91-172-2591407

Call Us : 08043692865

Digital Flame Photometer - 391 & 381


Digital Flame Photometer - 391 & 381


Digital Flame Photometer Model � 381 is a simple & easy to operate instrument for determination of Sodium, Potassium, Calcium & Lithium. It is an ideal instrument for Soil testing labs. Digital Clinical Flame Photometer Model - 391 is a Dual Channel Flame Photometer used for simultaneous observation of Sodium, Potassium in blood serum (plasma), Urine & other type of biological fluids.

The instruments are supplied in two units viz. Main Unit and Compressor Unit. The Main Unit comprises of digital displays, detector, filters and nebulizer. The compressor unit supplies the compressed air to the Main Unit

Air at a given pressure is passed into the atomiser. The suction produced by it draws the solution of the sample into the nebulizer and is sprayed as fine mist into the flame. The light emitted by the element at the characteristic wavelength is isolated by the narrow band interference filters, which then falls on to the silicon photodiode. After signal conditioning of the photodiode output, it is sent to digital converter and the results are displayed directly in ppm/meq value. The narrow band interference filters are mounted on a rotating disc fitted within the Main Unit. Appropriate filter can be selected by rotating this disc with the thumb. The oil free compressor with regulator maintains a stable airflow.

Features

  • Simple & Easy to Operate
  • Narrow Band Interference Filters
  • Accurate Na, K, Ca, & Li Determination
  • Dual Channel For Simultaneous Observation of Na & K with Single Aspiration (Model - 391)
  • Mixing Chamber Designed For Stable Flame
  • Concentric Non-Corrosive Nebulizer
  • Auto Ignition


Specifications

Model 381 391
    Serum Urine
Range Na: 0 - 100 ppm

K : 0 - 100 ppm


Ca: 15 - 100 ppm
(Optional)

Li: 10 - 100 ppm (Optional)
Na: 0 - 200 meq/l

K : 0 - 10 meq/l


-


Li: 0 - 2 meq/l (Optional)
Na: 0 - 200 meq/l

K : 0 - 100 meq/l (Optional)

Ca: 0 - 10 meq/l


-
Sensitivity Na: 5 ppm

K : 5 ppm

Ca: 10 ppm
(Optional)

Li: 10 ppm
(Optional)
-
Accuracy + 2% upto 40 ppm,
+ 5% above 40 ppm
+ 2% upto 100 meq
+ 5% above 100 meq
Measurement
System
One element
at a time
Two elements at a Time
Readout 2½ Digit
7-Segment LED
Dual 2½ Digit, 7-Segment LED
Ignition System In-built electronic Ignition by press of switch
Repeatability + 2 Counts
Detector Silicon Photodiode
Filters Narrow band interference glass filters
Nebulizer Black backelite, axial flow type
Flame System LPG & dry oil free air
Warm-Up Time 10 minutes
Power 230 V + 10% AC, 50 Hz
Dimensions 375 x 245 x 220 mm (L x B x H) (Approx.)
Weight 7.5 Kg (Approx.)
COMPRESSOR UNIT
Air Supply By oil free mini compressor unit with pressure regulator
Combustion Gas LPG controlled by precision regulator
Power 230 V + 10% AC, 50 Hz
Dimensions 290 x 255 x 210 mm (L x B x H) (Approx.)
Weight 8 Kg (Approx.)