Microprocessor Water & Soil Analysis Kit - 1160


Microprocessor Water & Soil Analysis Kit - 1160


For efficient use of nutrients in the soil, to test the purity of drinking water and for waste water testing, the company has introduced Microprocessor Water and Soil Analysis Kit – 1160. This is a unique portable instrument for measurement of various parameters i.e. pH, Conductivity, TDS, Salinity, Temperature, Dissolved Oxygen and mV of any solution. A chargeable sealed lead acid battery is provided within the kit that makes it extremely useful for field and laboratory use.

The use of latest microcontroller technology makes it reliable, accurate and low power consuming instrument. The instrument has an automatic decimal positioning, auto buffer recognition and polarity indication and low battery indication facility which eliminates manual errors.

Features
  • Measures pH, Conductivity, TDS, DO, Salinity, mV & Temperature
  • Highly Accurate & Stable
  • User Friendly & Menu Driven Software
  • 16 x 2 line Alphanumeric Backlit LCD Display
  • Printer Attachment facility
  • Storage for 90 Samples
  • Mains Cum Battery Operated


Download


Specifications

Paremeters pH Cond. TDS DO ORP Salinity Temp.
Range 0 to 14 pH 0-20, 0-200 uS, 0-2, 0-20, 0-200 ms 0-20, 0-200 ppm, 0-2, 0-20, 0-200 ppt 0-20 ppm 0 to ± 1999.9 mV 0-20, 0-200 ppm, 0-2,0-20, 0-50 ppt 0-1000C
Resolution 0.01 pH 0.01 pH 0.01 pH 0.1 mV 0.1 mV 0.01ppm 0.1oC
Accuracy 0.01 ± 1 Digit ± 0.5% FS ± 1Digit ± 0.5% FS ± 1Digit 0.1 ppm ± 1Digit ± 0.1 mV ± 1 Digit ± 1% of Range ± Digit 0.10 oC ± 1 Digi
Temp. Comp. 0 to 1000C 5-50oC (Auto/Manual) Auto/Manual Auto/Manual (5 to 550C)   Auto/Manual  
Sensor 0 to 100oC (Auto/Manual) Cond/Salinity/ TDS cell Cond/Salinity/ TDS cell Cond/Salinity/ TDS cell Platinum Cond/Salinity TDS cell Semi Conductor Type
Calibration Auto & Manual     Auto      
Reference Temperature   Selectable (5 - 50oC) Selectable (5-50oC)        
Auto Buffer Recog. 4.00, 7.00 & 9.2 pH            
Range   Automatic Automatic     Automatic  
Cell Constant   Adjustable (0.05 to 10.5) (0.05 to 10.5) (0.05 to 10.5)      

Power Supply : 6V Rechargeable Battery or 230 ± 10%AC, 50Hz.

 

Accessories : pH Electrode, Cond./TDS Cell, Temp. Probe, DO Probe, DO Membranes, Tablets 7 pH & 4 pH, Main Lead and Operation Manual.

Note : Specifications are subject to change due to continuous improvements.